MDM Observatory 1.3m Telescope Observer's Report for 2002 Oct 30 Observer(s): A. Stutz Institution(s): OSU Instrument: 8K Worked for 8 hours (first half of the night) Conditions were Photometric most of the night. Problems were encountered, see the separate trouble report for details. Seeing and Weather: Beg: Seeing 1.5 arcsec, Clear, Winds Calm Mid: Seeing 1.7 arcsec, Clear, Winds Calm End: Clear, Winds Calm Observing Summary: Observed fine untill 1:15AM, when the following problem occured: A set of four 300sec exposures was interupted and the following error message appeared in the detcom window (in red): TDL failed to get any response When I typed flush I got the following error: Can't communicate with 1st timing board; not downloading DSP code utility 2nd timing utility Need to have code downloaded and power on; clean failed Then I typed pof and I got a very similar error but it did turn the instrument off. Then I cycled power and closed the detcom window and restarted everything. When I restarted the detcom window I got the same errors listed above plus some new ones: wrmem with mask failed because rdmem failed WRM command to 1st utility board failed When I would hit enter I got this error: TDL failed to get any responses At this point I tried cycling power and restarting the detcom window various times, but it did not help. And when I typed pon the instrument did not turn on despite the error messages. I also checked the cables on the back of the instrument, to make sure they were correctly attached, which they were. ------------------------------ Submitted on 2002 Oct 31 [2:29:28]