MDM Trouble Report for 2001 Jun 30 Telescope: 2.4m Observer(s): John Thorstensen (Dartmouth) Instrument: modspec+echelle Problem(s) Encountered: This is not a serious problem but should be documented I think. About 1/3 of the time at the beginning of the night, the very first picture refuses to read. The error message is: Beginning CCD read of 332 by 1680 (557760 pixels)... ccd_aread: wrong aio finished; expected 0. This is bad. *** Error encountered in asynchronous read: data suspect Failed, return code = 0, elapsed = 2.557000 The cure is to df echelle and reset whatever format you use. Sometimes one has to wait a minute or two for the electronics to clear before the df echelle will take. I can easily imagine an inexperienced observer losing a fair amount of time to this. ------------------------------ Submitted on 2001 Jun 30 [19:33:35]