MDM Trouble Report for 2002 May 20 Telescope: 1.3m Observer(s): Fields & Burke (OSU) Instrument: Echelle Problem(s) Encountered: During readout of image following error message occured in CCDCOM: #2 Cleared chip 1 time Exposure started, ^C to pause Exposure complete, time 90.00 sec Beginning CCD read of 2080 by 2048 (4259840 pixels)... ccd_aread: wrong aio finished; expected 0. This is bad *** Error encountered in asynchronous read: data suspect Failed, return code = 0, elapsed = 1.189000 Issued following commands in CCDCOM df echelle sync ut init No more readout errors occured following this reinitialization. The same problem occured previous night and was fixed in the same manner. ------------------------------ Submitted on 2002 May 20 [5:24:26]