MDM Trouble Report for 2002 Oct 30 Telescope: 1.3m Observer(s): A. Stutz (OSU) Instrument: 8K Problem(s) Encountered: A set of four 300sec exposures was interupted and the following error message appeared in the detcom window (in red): TDL failed to get any response When I typed flush I got the following error: Can't communicate with 1st timing board; not downloading DSP code utility 2nd timing utility Need to have code downloaded and power on; clean failed Then I typed pof and I got a very similar error but it did turn the instrument off. Then I cycled power and closed the detcom window and restarted everything. When I restarted the detcom window I got the same errors listed above plus some new ones: wrmem with mask failed because rdmem failed WRM command to 1st utility board failed When I would hit enter I got this error: TDL failed to get any responses At this point I tried cycling power and restarting the detcom window various times, but it did not help. And when I typed pon the instrument did not turn on despite the error messages. I also checked the cables at the back of the instrument to make sure they were properly attached. ------------------------------ Submitted on 2002 Oct 31 [2:31:38]