MDM Trouble Report for 2003 Oct 17 Telescope: 1.3m Observer(s): Marshall, Pepper (OSU) Instrument: Templeton Problem(s) Encountered: The filter wheel seemed to seize up at about 2:30am. We tried fixing the problem by pointing the telescope to zenith, which allowed the wheel to move again but only very very slowly (it took about 2 minutes to move from one filter to the one next to it). For the rest of the night we will observe one target in only one filter, but if the filter wheel does not recover by tomorrow night this will present a real problem for our observing program which consists of BVRI imaging of many stars around the sky. I have also noticed that as we tried to continue to move the wheel--albeit very slowly--it responds more and more slowly, now taking >3 minutes to change between filters. I will not change filters again until I receive further instructions. I will try to call Bob in the morning to resolve the problem. ------------------------------ Submitted on 2003 Oct 18 [5:05:40]