MDM Trouble Report for 2006 Jul 31 Telescope: 2.4m Observer(s): Barr (MDM) Instrument: CCDS Problem(s) Encountered: Over the weekend the observers were having a problem with CCDS. When I tried rebooting the IC and the HE electronics, when the system was brought up on the mono display in it showed the -sequencer and -HE were not up and running. I used another computer and sequencer from the 1.3m and still had the same results. I used another set of fiber optic cable and still had same results. Following are the errors that were displayed on the mono screen for the IC: Error: Timeout waiting for word in from HE . (this is displayed many times) Error: No HE UARTS found. Disabling utility functions from sequencer Error: Trying to reset sequencer: Sequencer Acquisition disabled. Error Trying to reset head electronics: Head electronics utilities disabled. Error: Can't Initialized UARTS , Head electronics UARTs are disabled. So now I am awaiting instruction from Jerry et al ------------------------------ Submitted on 2006 Jul 31 [11:48:40]