MDM Trouble Report for 2006 May 17 Telescope: 2.4m Observer(s): Sebastien Lepine (AMNH) Instrument: Direct + Echelle Problem(s) Encountered: Around 1pm I got the following error from CCDCOM: *** Problem clearing CCD, Camera not responding I fixed it by doing several cl . But about 10 minutes later I got the following errors: *** WARNING: Did not receive shutter close indication from camera *** Problem clearing CCD, Camera not responding I then tried a hardware reset of the camera. It came back on, but on the next exposure I got this error message: *** Error encountered in asynchronous read: data suspect and no image would write down. I finally turned the gold power supply off and on again, apparently fixing the problem; I did not get any more errors afterward. ------------------------------ Submitted on 2006 May 18 [5:36:15]