MDM Trouble Report for 2007 Jun 1 Telescope: 2.4m Observer(s): Sivakoff (The Ohio State University) Instrument: Echelle Problem(s) Encountered: Ambient Dome Temperature Malfunctioned Twice: The first time was only seen in retrospect. The second time, I noticed a mirror close warning on the A/C monitor to a Dewpoint Warning. I closed the mirrors immediately. Review of the A/C history showed Ambient Dome Temperatures off the chart (this was when the first case was noticed). Since the MDM weather station, as well as several online forecasts, showed a more correct temperature and a low dewpoint, I ignored the warning and opened the mirror. At some point, the temperature sensor began operating normally by itself. Impact: 15 min 1 Software Error on Readout: Reading out a twilight flat, I got the following error message. Beginning CCD read of 2080 by 2048 (4259840 pixels)... ccd_aread: wrong aio finished; expected 2. This is bad. *** Error encountered in asynchronous read: data suspect Failed, return code = 1507328, elapsed = 44.152000 I tried clearing the CCD, but got *** Problem clearing CCD, Camera not responding. After a few tries, the camera began to respond again by itself. Impact: 5 min ------------------------------ Submitted on 2007 Jun 1 [5:45:40]