MDM Trouble Report for 2008 Nov 15 Telescope: 2.4m Observer(s): Desiree Cotto-Figueroa, David Riethmiller (Ohio University) Instrument: Echelle Problem(s) Encountered: Encountered a strange error with CCDCOM: 145> Clearing the CCD.... Done, cleared CCD 1 time Getting TCS/MIS info...done Exposure Started, Ctrl+C to pause Exposure Complete, Elapsed Time = 30.00 sec Beginning CCD read of 1232 by 1200 (1478400 pixels)... ccd_aread: wrong aio finished; expected 2. This is bad. *** Error encountered in asynchronous read: data suspect Failed, return code = 458752, elapsed = 17.273001 145> The error seems to have been bypassed by restarting the CCDCOM window. Also, we noticed that the hose on the liquid nitrogen tank is cracked in several places, allowing for small leaks during dewar filling. ------------------------------ Submitted on 2008 Nov 16 [6:51:42]