MDM Trouble Report for 2010 Jun 18 Telescope: 2.4m Observer(s): John Thorstensen (Dartmouth) Instrument: Modspec + Templeton CCD Problem(s) Encountered: At the start of the night, the Andor camera wouldn't come on line -- I'd left everything on during the day, but the Andor control program had evidently crashed and refused to restart, immediately reporting a timeout. MaximDL wouldn't connect to the camera either. This of course caused some anxious moments, but I solved it by cycling the power on the camera out at the telescope. There does not appear to be a power switch (the switch on the power brick toggles deep cooling and regular cooling), so I had to unplug the power cord to cycle the power. After this, the camera worked perfectly all night. This doesn't require any action, but I'm submitting it in the spirit of reporting problems so others may be made aware of the possibility. If this happens frequently we might want to turn the cameras off during the day. ------------------------------ Submitted on 2010 Jun 19 [5:07:51]