MDM Trouble Report for 2012 Feb 22 Telescope: 2.4m Observer(s): Kay et al. (Columbia) Instrument: Mk III + Nellie + 600/5800 Problem(s) Encountered: The instrument rotator wouldn't work at first. Could hear an interlock or something click when CW or CCW was pushed, but no rotation. We tried power cycling the rotator electronics box, which worked. We had to do this more than once. Lost a little time consulting the manual when it first happened, but no serious impact. We experimented with the lens position on the slit viewing camera and found that both the slit and objects looked sharper with the position a little beyond where Tony set it (so, even a little further from the Modspec position than the MkIII mark). The dewar tubing near the probe needs to be reclamped -- LN2 leaks out. ------------------------------ Submitted on 2012 Feb 23 [5:34:43]