MDM Trouble Report for 2013 Jan 11 Telescope: 1.3m Observer(s): R. Mark Wagner (OSU) Instrument: MDM4K Problem(s) Encountered: I would like to report unauthorized or spurious motion of the 1.3 m telescope. As I summarized in the night report for this night, the LN2 fill line snapped while filling the dewar. I found that the telescope had moved about 3 deg north in declination while filling which caused the snap. THis afternoon I filled the dewar with the interlock enabled and there was no problem. However, when I disabled the interlock the telescope started moving north in declination right away. The rate was about 2-3 arcmin per 10 seconds. Observers should exercise caution and be aware of this problem. Also, all LN2 fills should be done with the interlock enabled. ------------------------------ Submitted on 2013 Jan 13 [2:30:01]