MDM Trouble Report for 2015 Sep 04 Telescope: 2.4m Observer(s): Frank (OSU) Instrument: OSMOS Problem(s) Encountered: Changing the disperser, slit, or filter resulted very often in complaints about being out-of-position (every fifth or six times when a change had to be initiated). Luckily, the reset command always worked nicely, but the frequency of this happening is a bit worrisome. Below is an example of one of the error messages, resulting also in a bright red warning sign on the prospero status window at the appropriate place : Setting disperser... Error: Command returned error ERROR: DISPERSER DISPERSER=FAULT Mechanism out-of-position, in-position sensor not asserted. Reset DISPERSER to recover Error setting the Disperser Programming Error: invalid error code: -99 DISPERSER request aborted with errors. Execution error in Prospero command. Command: disp 2 ------------------------------ Submitted on 2015 Sep 5 [1:21:23]