MDM Trouble Report for 2016 Nov 24 Telescope: 2.4m Observer(s): Jaejin Shin (Seoul national University) Instrument: OSMOS Problem(s) Encountered: Around 2 am, background count was ~8000. I figured out it is because of high CCD temperature (~ -35). It is weird because I filled liquid nitrogen and 5:30pm and I checked the nitrogen leaking which means it is totally filled. Anyway, around 3 am, I refilled nitrogen again, temperature went down, and background count came back to normal value. ------------------------------ Submitted on 2016 Nov 25 [6:11:18]