MDM Trouble Report for 2018 Mar 01 Telescope: 2.4m Observer(s): John Thorstensen (Dartmouth) Instrument: modspec + echelle Problem(s) Encountered: The first spectrum of the night (of a bright star) came out looking like a bias, indicating that the shutter in modspec was not opening. I removed the shutter housing from the spectrograph and found that it would open when I operated it using the switch on the shutter power supply. Eventually found that the connection between the Leach controller and the shutter had failed because the connector at the power supply box had somehow fallen out. Eric told me the correct way to plug it back in (there were several possibilities) and after reassembling everything I was back on the air. I do hope that I didn't dislodge the O-ring between the shutter assembly and the spectrograph body in re-installing! Guide probe froze at the end of the night; I restarted it, but it'd be good to check it's OK. ------------------------------ Submitted on 2018 Mar 2 [6:30:43]