Templeton pattern noise
Tuesday, May 11, 2021 6:45 AM
Problem(s) Encountered:
Templeton's performance is degrading, in the form of pickup-like pattern noise with amplitude of several ADU, which is quite a few electrons. This stamps itself on faint spectra and makes it very hard to work on fainter objects -- you never know if that 20-min exposure is going to be any good. It's as bad as Echelle was getting, and appears to be a very similar issue. I doubt if much can be done about it, and I may be able to filter some of it out with enough effort, but I'd appreciate it if it's possible to have a look at it to see if anything obvious is wrong. I will try to remember to turn the controller off a the end of the night to see if resting it might help.
Solution:
Power line filter was swapped out for a new unit. Per observer: "The pattern noise is much improved, often absent and when present much less intrusive."