Revisiting when the dewar warms up

Monday, September 21, 2026 6:38 AM

Problem(s) Encountered:

In the morning, noticed that the instrument temperature increases to about -65 deg. After filling the dewar, the temperature drops again. I should have filled the dewar at about 17:45pm the night before. Hopefully it is just because I didn't fill the dewar full enough.

Solution:

It is better to overfill, allowing LN2 to freely pour from the overflow onto the floor than it is to underfill and risk the dewar warming up.  -65 is an appropriate temperature to go ahead and refill.  But if the dewar warms up beyond -50, it SHOULD not be refilled without consulting MDM staff.  Damage can occur if the detector is refilled after allowing it to go soft.